Thermo Scientific Evolution 60S UV-Visible Spectrophotometer
Built on the foundation of 60 years of experience in spectroscopy, the Thermo Scientific Evolution 60S UV-Visible spectrophotometer provides exceptional performance at an affordable price. Use the power of 1.0 nm resolution to take your measurements to the next level of clarity. Whether your application is research or routine analysis, you can count on the Evolution™ 60S for accurate and reliable data .
Evolution 60S Specifications
Optical Design | Dual Beam – Internal Reference Detector |
Spectral Bandwidth | 1.0 nm |
Light Source (Typical Lifetime) | Xenon Flash Lamp (5 years; 3 years guaranteed) |
Detectors | Dual Silicon Photodiodes |
Wavelength | |
- Range | 190 – 1100 nm |
- Accuracy | ± 0.8 nm |
- Repeatability | ± 0.5 nm |
- Slew Speed | 11,000 nm/min |
- Scan Speed | 10 – 4200 nm/min |
- Data Interval for Scanning | 0.1, 0.2, 0.5, 1.0, 2.0, 3.0, 5.0 nm |
Photometric | |
- Linear Range | > 3.5 A |
- Display | -0.5 – 5.0 A; -1.5 – 125% T; ± 9999 C |
- Accuracy | ± 0.005 A at 1.0 A 0.010 A K2Cr2O7 |
- Noise | < 0.00025 at 0.0 A, < 0.00050 at 1.0 A, < 0.00080 at 2.0 A, RMS at 260 nm |
- Drift | < 0.0005 A/hr |
Stray Light | < 0.08%T at 220, 340nm (NaI, NaNO2) < 1.0%T 198 – 200 nm (KCl) |
Resolution (Toluene in Hexane) | > 1.5 |
Display | Graphical with LCD backlight 9.7 x 7.1 cm (3.8 x 2.8 in.) |
Keypad | Sealed Membrane with tactile response keys |
Printer (optional) | 40 column Internal (text and graphics) External USB printer (HP PCL 3.0 and greater) |
Connectivity | USB Type A port for USB memory device (front panel) USB Type B port for optional computer connectivity (rear panel) USB Type A port for external printer (rear panel) |
Dimensions | 30 W x 40 D x 25 H cm (11.8 x 15.7 x 9.8 in) |
Weight | 8.6 kg (19 lb.) |
Power Requirements | Selected Automatically 100 – 240 V; 50 – 60 Hz |