Thermo Scientific Antaris II FT-NIR Analyzers
The Thermo Scientific™ Antaris™ II is a reliable FT-NIR analyzer that solves a broad range of industrial manufacturing quality control problems by providing timely measurements in the laboratory, at production point-of-use, or in real-time processes.
Antaris II System Specifications
Detection | High-sensitivity, high-stability matched InGaAs |
Interferometer | Proven frictionless, stable, long-life Michelson |
Instrument Dimensions | 40.6 cm (width) × 68.5 cm (depth) × 33 cm (height) |
Weight | 47.7 kg |
Source | Long-life, high intensity halogen NIR source; spare source included with system, guaranteed filament image alignment |
System Status Indicators | Indicator lights report scan, laser, power, and source status continuously |
Operator Communication Indicators | Red, yellow and green LED indicators communicate pass/fail/prompt |
Sealed and Desiccated | Yes |
Purge | Optional |
Operating Temperature Range | 15–35 °C |
Power Requirements | 90–264 VAC |
Integrated Computing | Optional |
Communications | Plug and Play USB communications to PC, no addressing or administration required |
Network and Control System Communications | Direct PC to Ethernet allows file system and OPC communications |
Regulatory Approvals | |
System Performance | |
Spectral Range | 12000–3800 cm-1 (833–2630 nm) |
Resolution | 4 cm-1 across spectral range (.6 nm at 1250 nm); 2 cm-1 option across spectral range (.3 nm at 1250 nm) |
Wavenumber Reproducibility (system to system) | Better than .05 cm-1 (.008 nm at 1250 nm) |
Wavenumber Repeatability (single system) | 10 measurement standard deviation <.006 cm-1 |
Wavenumber Accuracy | ±0.03 cm-1 (.005 nm at 1250 nm) |
Photometric Linearity (USP) | Slope 1.0 ± 0.05 and an intercept of 0.0 ± 0.05 |
Fiber Optic Module Specifications | |
Spectral Range | Range is dependent on fiber type and length of probe used |
Fiber Optic Connection | Standard SMA connection, allowing many probe sampling options |
SabIR Fiber Optic Probe Specifications | |
Spectral Range of SabIR Probe | 12000–4000 cm-1 (833–2500 nm) |
Probe Shaft | Stainless steel shaft 15.8 cm (length) × 1.6 cm (diameter) |
Fiber Optic Cable | High throughput low OH silica fiber bundle – two or three meter length |
Window Material | High quality, chemical-resistant sapphire |
Probe Weight | 0.7 kg |
Operating Temperature Range | 15–35 °C |
Remote Start Capabilities | Handheld trigger |
Background Handling | Automatic background in holster also available |
Operator communication indicators | Red, yellow and green LED indicators on probe communicate pass/fail/prompt |
Probe Holder | Holder with built-in Spectralon® reference |
Fiber Optic Connection | Standard SMA connection |
Rapid Liquid/Solid switching | Optional transflectance sleeves allow fixed or multiple pathlength settings, avoids primary probe window contamination |