Thermo Scientific Dionex ICS-6000 HPIC System
The Thermo Scientific™ Dionex™ ICS-6000 HPIC™ System is a high-pressure ion chromatography system that provides an impressive combination of increased productivity, expanded capabilities, and improved performance. Modular versatility, functional integration, superior performance and the capability to operate continuously at up to 5000 psi culminate in the most advanced ion chromatography system.
Pump | |
Type | Dual-piston (in series), microprocessor controlled, constant stroke, variable speed, patented Isokinetic Eluent Precompression |
Construction | Chemically inert, metal-free PEEK pump heads and flow path; compatible with aqueous eluents from pH 0–14 and reversed-phase solvents |
Pressure Range | Analytical: 0–35 MPa (0–5000 psi) Capillary: 0–41 MPa (0–6000 psi) |
Flow Rate Range | Analytical: 0.000–10.000 mL/min with settable flow increments at 0.001 mL/min, without changing pump heads. Optional heads allow flow up to 22.4 mL/min. Capillary: 0.001–0.100 mL/min flow rate and up to 3.000 mL/min for pump priming mL/min with settable flow increments at 0.0001 mL/min, typical working range of 5–30 μL/min |
Flow Rate Precision (Analytical and Capillary) | < 0.1% |
Flow Rate Accuracy | < 0.1% |
Pressure Ripple | < 1% at 1.0 mL/min/typical, < 0.2% (with damper) at 10 μL/min, < 1.0% (without damper) typical |
Pressure | None required |
Vacuum Degasser (Analytical and Capillary) | Integrated, optional 1 channel for isocratic pump or 4 channel for quaternary pump |
Piston Seal Wash | Standard, automatic operation |
Gradient Formation (Analytical and Capillary) | Quaternary low pressure (analytical only) or electrolytic eluent generation at high pressure |
RFIC-EG Gradients | Analytical: 0.1–100 mM Capillary: 0.1–200 mM |
Gradient Profiles | Any combination of an unlimited number of linear, convex, and concave positive and negative gradient profiles |
Gradient Proportioning Accuracy and Precision (Analytical Only) | ±0.5 at 2 mL/min |
Gradient Mixing (Analytical Only) | Passive mixers for 2 mm and 4 mm i.d. columns, optional |
Eluent Generation (Analytical and Capillary) | Optional eluent generation (RFIC-EG) |
Eluent On/Off Valve | Electrically actuated, standard |
Leak Sensor | Optical, standard |
System Software | |
Software | Chromeleon 7.2 CDS software, supports Microsoft® Windows® 7, operating systems |
Automated Procedure Wizards | Standard feature |
System Smart Startup and Shutdown | Standard feature |
Application Templates | Standard feature |
Automation Support of 3rd Party Instruments including | Fully controls over 300 different instruments from more than 30 manufacturers, GC, HPLC, and MS |
Customizable System Control Panels | Standard feature |
Signal Channels | Pump pressure |
Data Trending Plots | All numerical device parameters |
System Status Virtual Channels | Standard feature |
Power Failure Protection | Standard feature |
System Trigger Commands and Contitionals | Standard feature |
Daily Audit Trail | Standard feature |
Sample Audit Trail | Standard feature |
System Calibration Storage | Factory, present and previous. Completely user selectable |
Customized Reporting | Standard feature with unlimited report workbooks |
GLP Compliance | Optional |
Physical Specifications | |
Power Requirements | 90–265 V ac, 47–63 Hz |
Dimensions (h × w × d) | 41 × 23 × 56 cm (16 × 8.75 × 21.5 in.) |
Weight | SP: 20.4 kg (45 lb) — DP: 24.1 (55 lb) |
Flow Path: | All polymeric (PEEK), anion or cation configuration |
Minimum and Maximum Eluent Concentration | Analytical: 0.1–100 mM Capillary: 0.1–200 mM |
Flow Rates | Analytical: 0.1–3.000 mL/min Capillary: 0.001–0.030 mL/min |
Eluent Types | Analytical: KOH, LiOH, NaOH; Carbonate; Carbonate/Bicarbonate; MSA Capillary: KOH, MSA |
Maximum Operating Pressure | Analytical: Dionex EGC III: 21 MPa (3000 psi); Dionex EGC 500: 35 MPa (5000 psi) Capillary: 35 MPa (5000 psi) |
Maximum Solvent Concentration (Analytical and Capillary) | Cations: None Anions: 25% methanol (KOH, NaOH and LiOH cartridges); None (Carbonate cartridge and EPM) |
Gradient Profiles | Standard - any combination of an unlimited number of linear, convex and concave positive and negative gradient profiles |
Number of Cartridges Supported | Two–dual cartridge support RFIC Eluent Degasser (Analytical only) Eluent degasser housed in the EG Module |
System Software | |
Software | Chromeleon 7.2 CDS software, supports Microsoft Windows 7 operating systems |
Automated Procedure Wizards | Standard feature |
System Smart Startup and Shutdown | Standard feature |
Application Templates | Standard feature |
Automation Support of 3rd Party Instruments | Fully controls over 300 different instruments from more than 30 manufacturers, including GC, HPLC, and MS |
Customizable System Control Panels | Standard feature |
Signal Channels | Eluent concentration |
Data Trending Plots | Device numerical parameters plotted |
System Status Virtual Channels | Standard feature |
Power Failure Protection | Standard feature |
System Trigger Commands and Contitionals | Standard feature |
Daily Audit Trail | Standard feature |
Sample Audit Trail | Standard feature |
Eluent Cartridge Information Storage | Serial number and expiration date |
Customized Reporting | Standard feature with unlimited report work |
GLP Compliance | Optional package provides security, modification history, and electronic signatures |
Physical Specifications | |
Power Requirements | 90–265 V ac, 47–63 Hz |
Dimensions (h × w × d) | 41 × 23 × 56 cm (16 × 8.75 × 21.5 in.) |
Weight | 25 kg (40 lb) without optional items |
Flow Path | All polymeric (PEEK), anion or cation configuration |
Eluents | Carbonate and carbonate/bicarbonate combinations up to 20 mM MSA up to 34 mM |
Flow Rates | 1.00–2.00 mL/min |
Continuous Operation with 4 L of Eluent | Up to 28 days or 2000 samples, typically |
Always on, Always Ready Capable | Standard feature |
Remains Fully Calibrated for Extended Periods of Time (≤ 28 Days): | Standard feature, results are traceable to a single calibration |
System Wellness | Consumables usage monitoring for predictive maintenance |
Maximum Operating Pressure | 21 MPa (3000 psi) |
Operating Temperature Range | 4–40 °C |
Standard DC Model: | Upper Zone Temp Range: 18–40 °C (minimum temperature: ambient -15 °C); (maximum temperature: ambient +20 °C) Lower Zone Temp. Range: 10–70 °C (minimum temperature: ambient -15 °C) (maximum temperature: ambient +50 °C) Temperature Accuracy: ±0.15 °C Temperature Stability: <0.05 °C Temperature Precision: ±0.2 °C |
Low Temperature DC Model | Upper Zone Temp. Range: 10–40 °C (minimum temperature: ambient -17 °C); (maximum temperature: ambient +20 °C) Lower Zone Temp. Range: 10–70 °C (minimum temperature: ambient -15 °C or upper zone -20 C) (maximum temperature: ambient +50 °C) Temperature Accuracy: ± 0.15 °C Temperature Stability: <0.05 °C Temperature Precision: +/-0.2 °C |
Lower Zone (Analytical) | Injection Valves: Up to two high-pressure valves,either 6- or 10-port, 2-position Low Pressure Valves: Up to 2 inert, 2- or 3-way Reaction Coil Heater (RCH): Holds two reaction coils RCH Temperature Range: 5 °C above upper zone, 80 °C maximum AutoPrep Kit: Dual loop for sample preconcentration All valves and heater upgradable and field installable |
Automation Manager (Optional – Analytical Only capillary chemistries) | Injection Valves: Up to two high-pressure valves, either 6- or 10-port, 2-position Low Pressure Valves: Up to 2 inert, 2- or 3-way Reaction Coil Heater (RCH): Holds two reaction coils RCH Temperature Range: 5 °C above upper zone, 80 °C maximum AutoPrep Kit: Dual loop for sample preconcentration All valves and heater upgradable and field installable |
Capillary Dionex IC Cube Module (Optional – but required for Injection) | One or two Dionex IC Cube models holding the folllowing: |
Injection Valves: Up to two (one per cube( high-pressure valves, 4-port, 2-position. An optional 6-port, 2-position valve is available to support large loop and concentrator applications EG Degasser Column and Guard Carbonate Removal Device Capillary Suppressor (See Suppressor section) Dionex IC Cube modules and cartridges are customer installable with pre-formed tubing and color-coded labeling Capillary column heater 15 to 80 °C) |
|
Flow Path | Plumbing configurations for 4 mm, 2 mm, and 0.4 mm columns, fully inert, PEEK |
Detectors | Any combination of two conductivity or electrochemical detectors, upgradeable and field installable. Dual detectors can be operated simultaneously or independently. Optional remote operation up to 3 m from instrument. |
Analog Signal Output | Two analog output channels, two 24 V relays, 2 TTL Out and 8 TTL In lines, upgradable and field installable, optional |
Leak Detection | Optical leak sensor standard |
Application Control Automation | Standard. All DC modules come equipped with two switched AC controllers built-in to automate external devices and the ability to control up to six low pressure 2-way or 3-way valves for fluidic automation |
Suppression | |
Non-Suppressed Conductivity | Supported - both anion and cation |
Suppressor Wear Parts | None. No valves, pumps, peristaltic pump tubing, or inline filters required |
Chemical Suppression | 2 mm and 4 mm anion and cation membrane suppression available |
Displacement Chemical Suppression | 2 mm and 4 mm anion and cation membrane suppression available |
Electrolytic Suppression, Recycle Mode External Water Mode | 0.4 mm (with optional Dionex IC Cube module) 2 mm and 4 mm anion and cation. Micro-membrane (2 mm and 4 mm) or capillary tube membrane (0.4 mm). |
Salt Converter | Available in 2 and 4 mm versions |
Dionex ACRS-ICE | Available in 9 mm and 4 mm version |
Carbonic Acid Removal for Anions | Thermo Scientific™ Dionex™ ADRS 600 Anion Dynamically Regenerated Suppressor or Thermo Scientific™ Dionex™ ACRS™ Anion Chemically Regenerated Suppressor plus Thermo Scientific Dionex CRD 200 or Dionex CRD 300 Carbonate Removal Device for 2 and 4 mm versions, Therm Scientific™ Dionex™ ACES™ 300 Anion Capillary Electrolytic Suppressor plus Dionex CRD 200 Carbonate Removal Device (0.4 mm) for capillary suppressors |
Suppressors | |
Suppression Capacities | Dionex ADRS 600 (4 mm) 200 μeq/min Dionex ADRS 600 (2mm) 50 μeq/min Dionex AERS 500 Carbonate (4 mm) 30 μeq/min . Dionex AERS 500 Carbonate (2 mm) 7.5 μeq/min Dionex AERS 500e (4 mm) 200 μeq/min Dionex AERS 500e (2 mm) 50 μeq/min Dionex CDRS 600 (4 mm) 100 μeq/min Dionex CDRS 600 (2 mm) 35 μeq/min Dionex CERS 500e (4 mm) 100 μeq/min Dionex CERS 500e (2 mm) 35 μeq/min Dionex ACES 2 μeq/min Dionex CCES 1.5 μeq/min Dionex ACRS (4 mm) 150 μeq/min Dionex ACRS (2 mm) 37.5 μeq/min Dionex CCRS (4 mm) 150 μeq/min Dionex CCRS (2 mm) 37.5 μeq/min Dionex ACES 300 (0.4mm) 2 μeq/min Dionex CCES 300 (0.4 mm) 1.5 μeq/min |
Suppressor Void Volumes | Dionex ADRS 600 (4 mm) < 50 μL Dionex ADRS 600 (2mm) < 15 μL Dionex AERS 500 Carbonate (4 mm) < 50 μL Dionex AERS 500 Carbonate (2 mm) < 15 μL Dionex AERS 500e (4 mm) < 50 μL Dionex AERS 500e (2 mm) < 15 μL Dionex CDRS 600 (4 mm) < 50 μL Dionex CDRS 600 (2 mm) < 15 μL Dionex CERS 500e (4 mm) < 50 μL Dionex CERS 500e (2 mm) < 15 μL |
Software | Chromeleon 7.2 CDS software, supports Microsoft Windows operating systems |
Automated Procedure Wizards | 2 mm and 4 mm anion and cation membrane suppression available |
Electrolytic Suppression, Self-Regenerating | Standard feature |
System Smart Startup and Shutdown | Standard feature |
Application Templates | Standard feature |
Automation Support of 3rd Party Instruments | Fully controls over 300 different instruments from more than 30 manufacturers, including GC, HPLC, and MS |
Customizable System Control Panels | Standard feature |
Signal Channels | Detector signals, detector background signals, temperatures |
Data Trending Plots | All device numerical parameters plotted |
System Status Virtual Channels | Standard feature |
Power Failure Protection | Standard feature |
System Trigger Commands and Conditionals | Standard feature |
Daily Audit Trail | Standard feature |
Sample Audit Trail | Standard feature |
System Calibration Storage | Factory, present, and previous. Completely user selectable |
GLP Compliance | Optional |
Physical Specifications | |
Power Requirements | 90–265 V ac, 47–63 Hz; (auto-sensing power supply; no manual voltage or frequency adjustment required) |
Dimensions (h × w × d) | Standard Model 44.5 × 42 × 57.5 cm (17.5 × 16.0 × 22.6 in.) Low Temp Model 44.5 × 42 × 59.7 cm (17.5 × 16.0 × 23.5 in.) |
Weight | Standard Model 38 kg (84 lb) Low Temp Model 43 Kg (94 lb) |
Flow Path | All polymeric (PEEK), anion or cation configuration |
Electronics Type | Microprocessor controlled digital signal processing |
Cell Drive | 8 kHz square wave |
Linearity | R2≥0.999% |
Resolution | 0.00238 nS/cm |
Flow Path | Plumbing configurations for 4 mm, 2 mm, and 0.4 mm columns, fully inert, PEEK |
Output Range | Digital Signal Range: 0–18,000 μS/cm |
Analog Signal Range: 0–18,000 μS/cm | |
Noise, Wet | < 0.2 nS at 23 μS/cm background |
< 0.1 nS at 1 μS/cm background | |
Filter | Rise times 0 to 10 s, programmable |
Sample Rate | 1 to 100 Hz, user settable or automatic |
Cell Temperature | 5 °C above DC upper zone temperature to 60 °C maximum. User settable, working range is identical to settable range. |
Cell Temperature Stability | < 0.001 °C |
Cell Temperature Compensation | Default 1.7% per °C; programmable from 0–3% per °C |
Flow Cell Maximum Pressure | 10 MPa (1500 psi) |
Flow Cell Volume | Analytical: 0.7 μL Capillary: 0.02 μL |
Cell Electrodes Cell Body Heat Exchanger | Passivated 316 stainless steel. Compatible with MSA Chemically inert polymeric material Inert, tortuous-path for low axial dispersion |
Software | |
Software | Chromeleon 7.2 CDS software, supports Microsoft Windows 7, operating systems |
Automated Procedure Wizards | Standard feature |
System Smart Startup and Shutdown | Standard feature |
System Wellness and Predictive Performance | Standard feature |
Application Templates | Standard feature |
Automation Support of 3rd Party Instruments | Fully controls over 300 different instruments from more than 30 manufacturers, including GC, HPLC, and MS |
Customizable System Control Panels | Standard feature |
Signal Channels | Electrochemical and total electrochemical signals |
Data Trending Plots | All detector numerical parameters plotted |
System Status Virtual Channels | Standard feature |
Power Failure Protection | Standard feature |
System Trigger Commands Standard feature and Conditionals | Standard feature |
Daily Audit Trail | Standard feature |
Sample Audit Trail | Standard feature |
System Calibration Storage | Factory, present, and previous. Completely user selectable. |
Customized Reporting | Standard feature with unlimited report workbooks |
GLP Compliance | Optional |
Physical Specifications | |
Dimensions (h × w × d) | 6.9 × 16.7 × 9.9 cm (2.7 × 6.5 × 3.9 in.) |
Weight | 400 g (1.6 lb) |
Electronics Type | Microprocessor controlled digital signal processing |
Electronic Noise (Wet Noise) (Analytical and Capillary) | IPAD (Au electrode) < 30 pC at 10 mM KOH, DC Amperometry (GC) < 5 pA at catecholamine eluent |
Potential Range | -2.0 to 2.0 V in 0.001 V increments |
Signal Range (Digital and Analog) | Integrated Amperometry: 0.0008 pC to 200 μC DC Amperometry: 0.00008 pA to 74 μA |
Filter | 0–10 s response time, user settable |
Control Mode | Local or remote control using relay closures or TTL, or control using Chromeleon CDS software via DC module |
Cell Body | Titanium body with narrow-bore PEEK inlet tubing for capillary format, titanium inlettubing for analytical format. Compatible with 0.2–0.6 mm i.d. columns (PEEK inlet), 2–7 mm i.d. columns (Ti inlet) |
Working Electrodes | Conventional: gold, glassy carbon, platinum, and silver Disposable: gold, platinum, carbon, and silver |
Reference Electrode | pH-Ag/AgCl combination, one piece design (Analytical and Capillary) PdH, one piece design (Capillary Only) |
Autoranging | Yes |
Analog Output | User selectable full scale of 10, 100, or 1000 mV |
Cell Volume at Working Electrode | < 0.2 μL |
Maximum Cell Operating Pressure | 0.7 MPa (100 psi) |
Software | |
Software | Chromeleon 7.2 CDS software, supports Microsoft Windows 7, operating systems |
Detection Modes | DC amperometry, pulsed amperometry, and integrated amperometry. Detection modes include use of multiple waveforms and multiple integration times to optimize detection conditions for individual analytes. Integrated amperometry mode—unlimited changes to the waveform profile’s number of segments, duration of each segment, and voltage applied at each segment. |
Automated Procedure Wizards | Standard feature |
System Smart Startup and Shutdown | Standard feature |
System Wellness and Predictive Performance | Standard feature |
Application Templates | Standard feature |
Automation Support of 3rd Party Instruments | Fully controls over 300 different instruments from more than 30 manufacturers, including GC, HPLC, and MS |
Customizable System Control Panels | Standard feature |
Signal Channels | Electrochemical and total electrochemical signals |
Data Trending Plots | All detector numerical parameters plotted |
System Status Virtual Channels | Standard feature |
Power Failure Protection | Standard feature |
System Trigger Commands and Conditionals | Standard feature |
Daily Audit Trail | Standard feature |
Sample Audit Trail | Standard feature |
System Calibration Storage | Factory, present, and previous. Completely user selectable. |
Customized Reporting | Standard feature with unlimited report workbooks |
GLP Compliance | Optional |
Physical Specifications | |
Dimensions (h × w × d) | 6.9 × 16.7 × 9.9 cm (2.7 × 6.5 × 3.9 in.) |
Weight | 400 g (1.6 lb) |
Optics | |
Photodiode Array | 1024 element |
Pixel Resolution | 0.7 nm |
Lamps | Tungsten and deuterium |
Optical Resolution | 1.0 nm |
Wavelength Range | 190–800 nm |
Electronics | |
Analog Outputs | Four, 0–3 AU, 1000 mV range |
Control Modes | Software remote control through Chromeleon CDS software |
Flow Cell | |
Standard | PEEK or SST, 13 μL, 10 mm path length |
Semi-Prep | PEEK, 0.7 μL, 0.4 mm path length |
Maximum Flow Operating Pressure | 300 psi (< 2 MPa) PEEK 500 psi (< 3 MPa) SST |
Performance | |
Noise | ±10 μAU at 254 nm (flowing water, 2s rise time) ±15 μAU at 520 nm (flowing water, 2s rise time) |
Drift | < 500 μAU/h |
Wavelength Accuracy | ±1 nm, self-calibration with deuterium lines, verification with built-in holmium oxide filter |
Linearity | > 2 AU |
Physical Specifications | |
Power Requirements | 90–265 V ac, 47–63 Hz |
Operating Temperature Range | 4–40 °C (40–104 °F) |
Operating Humidity Range | 5–95% relative, noncondensing |
Dimensions (h × w × d) | 17.4 × 44.4 × 50.3 cm (6.8 × 17.5 × 19.8 in.) |
Weight | 18.1 kg (40 lb) |
Optics | |
Optical System wavelength UV-vis | Dual beam forward optics design (monochromator) single or multiple (4-channel) detector options |
Light Source | Deuterium lamp for UV range; tungsten lamp for visible range |
Wavelength Range | 190 to 900 nm in 1 nm increments |
Electronics | |
Analog Output Ranges (optional) | 0.001–3.0 AU |
Rise Time | 0.00–4.55 s |
Full-Scale Recorder Output | 1 V or 10 V |
Data Collection rate up to 100 Hz Digital Output | Full dynamic autoranging digital absorbance signal output to Chromeleon CDS software |
Control Modes | Software remote control through Chromeleon CDS software |
Flow Cells | |
Cell | PEEK |
Cell Path Length | 10 mm (Standard and Capillary), 7 mm (Micro) |
Cell Volume | 11 μL (Standard) 2.5 μL (Micro) 0.18 μL (Capillary) |
Heat Exchanger Volume | 8.8 μL |
Maximum Flow Cell Operating Pressure | 725 psi (5 MPa) |
Performance | |
Wavelength Accuracy | ±1 nm |
Linearity | > 2 AU |
Bandwidth | 6 nm at 254 nm |
Noise | Typically - < ±2.5 × μAU at 254 nm |
Drift | < 0.1 m μAU/h at 254 nm |
Physical Specifications | |
Power Requirements | 85–265 V ac, 47–63 Hz |
Operating Temperature Range | 10–35 °C (50–95 °F) |
Operating Humidity Range | 5–80% relative, noncondensing |
Dimensions (h × w × d) | 15.2 × 44.4 × 50.3 cm (6.0 × 17.5 × 19.8 in.) |
Weight | 15.4 kg (34 lb) |
Sample Capacity | 10 mL vials: 81 1.5 mL vials: 120 0.3 mL vials: 120 Well Plates: 288 (3 × 96 standard or deep well) Well Plates: 1152 (3 × 384 standard well) |
Minimum Sample Volume | 10 μL can be sampled from a 300 μL microvial; 20 μL can be sampled from a 500 μL microvial |
Maximum Injection Volume | 7500 μL |
Variable Volume Range | 1–100 μL in 0.1 μL increments; 100–7500 μL in 1 μL increments |
Injection Cycle Time (including sample prep such as dilution) | 15 s with sample overlap function; 30 s for a 5 μL full loop without sample overlap |
Injection Modes | Full Loop; Partial Loop; Limited Sample; Capillary |
Sampling Technique | Pull Mode; Push Mode |
Injection Precision | Fixed Loop: < 0.3% RSD at 20 μL; Partial Loop: < 0.5% RSD at 20 μL Capillary Mode: < 0.5% RSD at 0.4 μL |
Dilution | 1:1 to 1:1000 |
AutoDilution | With post-run functionality and Chromeleon AutoDilution License |
Dilution Precision | < 1.0% RSD for a 1:10 dilution |
Carryover | < 0.01% with 500 μL flush volume |
Sample Tray Thermostatting | 4 °C to 60 °C, optional |
Sample Conductivity and pH | Inline measurement of sample conductivity and pH with postrun functionality, optional |
Dual Instrument Support | Sequential (asynchronous) and Simultaneous modes, field upgradeable, optional |
Sample Degassing | CRD 200/300, upgradeable, user installable |
Injection Valves | One or two 2-position, 6- or 10-port |
Diverter Valve | One 2-position, 6- or 10-port |
Sample Preparation | Dilution, addition of internal standards, concentration/matrix elimination, derivatization, etc. |
Inline Sample Filtration | Dual filter, backflush, optional |
Software | |
Software | Chromeleon 7.2 CDS software, supports Microsoft Windows 7, operating systems |
Automated Procedure Wizards | Standard feature |
System Smart Startup and Shutdown | Standard feature |
System Wellness and Predictive Performance | Standard feature |
Application Templates | Standard feature |
Automation Support of 3rd Party Instruments | Fully controls over 300 different instruments from more than 30 manufacturers, including GC, HPLC, and MS |
Customizable System Control Panels | Standard feature |
Power Failure Protection | Standard feature |
System Trigger Commands and Conditionals | Standard feature |
Daily Audit Trail | Standard feature |
Sample Audit Trail | Standard feature |
System Calibration Storage | Factory, present, and previous. Completely user selectable. |
Customized Reporting | Standard feature with unlimited report workbooks |
GLP Compliance | Optional |
Physical Specifications | |
Dimensions (h × w × d) | 44.5 × 35.9 × 51.6 cm (17.5 × 14.1 × 20.3 in.) |
Weight | < 25.2 kg (< 55.5 lb) < 26.8 kg (< 59.0 lb) with sample temperature control |
Power | 90–265 V ac, 47–63 Hz |
Flow Path | All polymeric (PEEK), anion or cation configurations |