UltiMate 3000 series

Thermo Scientific UltiMate 3000 RSLCnano UHPLC System

The Thermo Scientific™ UltiMate™ 3000 RSLCnano modular system offers precision, productivity and
performance for all your low flow UHPLC needs

Thermo Scientific UltiMate 3000 RSLCnano System
Recommended flow range50 nL/min–50 μL/min (controlled with dedicated flow meters); 50–2500 μL/min with ternary micro pump
(only in NCS-3500RS)
Pressure rangeWith nano ProFlow flow meter: 860 bar (12,500 psi)
With capillary flow meter: 800 bar (11,600 psi)
With micro flow meter: 800 bar (11,600 psi)
System delay volume<350 nL in pre-concentration configuration
Retention time RSD≤0.2% RSD or < 0.1 SD whichever is greater with a 30 min gradient
Safety featuresSystem wellness monitoring, leak sensors, active rear-seal wash system, excess pressure monitoring,
emergency shut-down
pH range2–10
Wetted partsMP35N, titanium, PEEK, UHMW-PE, PTFE, FEP, sapphire, ZrO2, Al2O3, fused silica, SST, PCTFE, perfluoro
elastomer (FFKM)
NCS-3500RS - Binary High Pressure Gradient (HPG) Pump with Ternary Micro Pump and Column Compartment
NCP-3200RS - Binary HPG Pump
Binary High Pressure Gradient (HPG) Pump
Flow rate range and maximum pressureNano ProFlow flow meter: 0–1,500 nL/min (recommended: 50–1,500 nL/min)
900 bar (13,050 psi) at full flow range
Capillary flow meter: 0–15 μL/min (recommended: 0.5–10 μL/min)
800 bar (11,600 psi) at nominal flow rate 5 μL/min
Micro flow meter: 0–50 μL/min (recommended: 5–50 μL/min)
800 bar (11,600 psi) at nominal flow rate 25 μL/min
Gradient delay volume< 25 nL
Number of solvent channels2
Wetted parts HPGMP35N, titanium, PEEK, UHMW-PE, PTFE, FEP, ZrO2, Al2O3, fused silica, sapphire, PCTFE, SST
Ternary Micro Pump (Only in NCS-3500RS)
Micro pump (low pressure gradient)Flow rate range: 0–2500 μL/min; recommended 5–2500 μL/min; gradient formation is recommended
from 50 μL/min
Maximum pressure620 bar (9,000 psi)
Number of solvent channels3
Delay volume220 μL
Proportioning accuracy/precision±1.0% of full scale/<0.3% SD
Wetted partsTitanium, PEEK, UHMW-PE, PTFE, FEP, ZrO2, Al2O3, perfluoro elastomer (FFKM)
Column Compartment (Only in NCS-3500RS)
Temperature rangeRoom temperature + 7 ˚C up to 75 ˚C
Temperature accuracy/precision±0.5 ˚C (at 50 ˚C setpoint)/±0.1 ˚C
Temperature stability±0.1 ˚C (at 50 ˚C setpoint)
Heat-up timeFrom 35 ˚C to 65 ˚C in 12 min at an ambient temperature of 25 °C
Switching valvesUp to two valves
6-port 2-position (port-to-port volume: 91 nL, maximum pressure: 860 bar (12,500 psi))
10-port, 2-position (port-to-port volume: 114 nL, maximum pressure: 860 bar (12,500 psi))
CapacityMaximum available width for column plus fittings: 350 mm; coiled fused silica columns
FeaturesHumidity sensor (column compartment), leak sensor, gas leak sensor, active rear-seal wash system,
excess pressure monitoring
NCS-3500RS/NCP-3200RS Features and Dimensions
GLP featuresColumn tracking
DimensionsNCS-3500RS (h × w × d): 36 × 42 × 51 cm (14.2 × 16.5 × 20 in.)
NCP-3200RS (h × w × d): 21 × 42 × 51 cm (8.3 × 16.5 × 20 in.)
WeightNCS-3500RS: 32 kg (70.6 lb); NCP-3200RS: 17.5 kg (38.6 lb)
Power requirements (automatic voltage selection)100–120 V, 60 Hz
200–240 V, 50 Hz; max 300 VA
PC connectionUSB 2.0; USB hub with three integrated sockets
I/O interfacesTwo digital inputs and two programmable outputs
Additional communication port15-pin D-Sub port for connection of a solvent rack or degasser
Autosampler (WPS-3000TPL RS)
Injection volume range0–20 μL (recommended 20 nL–20 μL with different nanoViper sample loops); upgrade kit for injection
volumes up to 125 μL available
Sample capacity3 × well plates (128 × 86 mm)
15 × 10 mL vials for reagents, diluents, and transport liquids*
Sample formats96 (deep) well plate, 384 (deep) well plate, sealed or open;
40 standard autosampler vials, 1.8 mL, sealed or open*
Injection cycle time<30 s for a 1 μL full-loop injection
Maximum pressure860 bar (12,500 psi)
Injection methodsFull loop and partial loop injections, low-dispersion mode, μL pick-up, user-defined programs
Injection techniqueNeedle-in-needle with programmable needle wash
Injection precision<0.4% RSD for 1 μL full loop injection
Injection linearityCorrelation coefficient > 0.9995, at 100 to 500 nL partial-loop injections, caffeine in water
Carryover<0.02% for caffeine with external wash
Sample cooling4–45 ˚C, or 22 ˚C below ambient
Biocompatible versionYes (upgrade kit available);
Fraction collectionYes, with sample cooling (WPS-3000TFC, extra parts required: nano/cap modification kit and 6-port
2-position C72 valve)
Wetted partsPEEK, SST, PAEK, PCTFE, PEEKsil™, fused silica
Dimensions (h × w × d)36 × 42 × 51 cm (14.2 × 16.5 × 20 in.)
Weight24 kg (53 lb)
Power requirements (automatic voltage selection)100–120 V, 60 Hz
200–240 V, 50 Hz
PC connectionUSB; USB hub with three integrated sockets
I/O interfacesFour digital inputs and four programmable outputs
UV/Vis Detector (VWD-3400RS)
Data collection rateUp to 200 Hz (in single wavelength mode); 5 Hz in multichannel mode
Maximum number of channels4
LampsDeuterium lamp, Tungsten lamp
Temperature control for both lamps
Drift4.0 mAU/h
Wavelength range190–900 nm ±1 nm
NoiseTypically < 0.05 mAU at 254 nm
Wavelength accuracy1 nm
Flow cell volume3 nL for nano LC; 45 nL for capillary LC; 180 nL for micro LC
Dimensions (h × w × d)16 × 42 × 51 cm (6.3 × 16.5 × 20 in.)
Weight15 kg (33 lb)
Power Requirements (automatic voltage selection)100–120 V, 60 Hz
200–240 V, 50 Hz
PC connectionUSB
I/O and analog interfacesTwo digital inputs and two programmable outputs
Two analog inputs available as an option via DAC plug-in module
Solvent Rack with Degasser (SRD-3400)
CapacitySixteen 0.5 L reservoirs
Degassing channels4 (SRD-3400); 0 (SR-3000)
Channel volume670 μL
Wetted materialsTeflon AF, PEEK, and Tefzel
Status LEDsPower, vacuum pump status, error (vacuum and/or leak)
Power supply and communication15-pin D-Sub (through UltiMate 3000 pumps) or external power supply
Dimensions (h × w × d)10 × 42 × 51 cm (3.9 × 16.5 × 20 in.)
WeightWith internal degasser: 4.8 kg (10.6 lb)
*Contact local sales representative for specific information
Software
Stand-alone UltiMate 3000 RSLCnano systemThermo Scientific™ Chromeleon™ 7.2 SR4 or later, Chromeleon 6.8 SR16 or later*
UltiMate 3000 RSLCnano with Thermo Scientific MS detectorsThermo Scientific Standard Instrument Integration (SII) 1.2 or later for Xcalibur™ and Chromeleon 7.2 SR4 or later; Chromeleon 7.2 SR4 or later*
UltiMate 3000 RSLCnano with Bruker MS detectorsDCMSLink™ based on Chromeleon 6.8 SR16 or later for Bruker Compass™/HyStar™*
*Contact local sales representative for more details about instrument control

Product Information PDF

UltiMate 3000 series

Q Exactive Orbitrap GCMS

Thermo Scientific Orbitrap GC-MS Contaminants Library

Using the Thermo Scientific™ Orbitrap™ GC-MS systems for the analysis of pesticides and other food and environmental contaminants has unique advantages over alternative techniques. This technology allows analytical chemists to achieve comparable sensitivity, selectivity and dynamic range performance to GC triple quadrupole technology using an untargeted full-scan acquisition, enabling:
  • The targeting of an unlimited number of analytes down to triple quadrupole level detection limits
  • Identification of unknowns driven by spectral matching and sub-ppm mass accuracies
  • Retrospective screening of compounds not known to be of interest at time of acquisition
  • Product Information PDF

    Q Exactive Orbitrap GCMS

    TRACE-1300 & 1310-GC

    Thermo Scientific TRACE 1300 Series Gas Chromatograph

    The Thermo Scientific™ TRACE™ 1300 Series Gas Chromatograph is the latest technology breakthrough conceived to substantially elevate performance in QA/QC and routine laboratories. Engineered around newly developed proprietary injectors and detectors, available as user-exchangeable, instant connect modules, these GC platforms offer greater flexibility over previous instrumentation and drastically reduce the cost of ownership.

    Performance Specifications
    • Typical retention time repeatability : <0.0008 min
    • Typical peak area repeatability : <0.5 % RSD
    Oven Specifications
    • Column oven (H × W × D) : 27 × 27 × 17.7 cm; 12.9 L
    • Operating temperature range : ambient +3 °C to 450 °C
    • Cryogenic option minimum temperature : -100 °C with liquid nitrogen; -50 with liquid CO2
    • Temperature set point resolution : 0.1 °C
    • Number of ramps/plateaus : 32/33
    • Maximum heating rate : 125 °C/min
    • Oven cool-down (22 °C ambient) : 450 °C to 50 °C in less than 4 minutes
    • Ambient rejection : < 0.01 °C per 1 °C

    Product Information PDF

    TRACE-1300 & 1310-GC

    Trace 1300 Series Catalogue

    The productivity you need the performance you want

    Reducing cost of ownership while increasing productivity is one of today’s tough challenges faced by
    laboratories using automated analytical equipment. For QA/QC and routine chemical laboratories, improvements in this area are intimately connected with the simplification of analytical workflows, optimization of technical resources and waste reduction.
    The Thermo Scientific TRACE 1300 Series Gas Chromatograph is the latest technology breakthrough conceived to substantially elevate performance in QA/QC and routine laboratories. Developed around key innovations driven by customer needs, including user-exchangeable instant connect injectors and detectors, enhanced robustness of components, scaling down of injector and detector solutions, customer-driven design of the user interface, and optimization of all electronic elements, the TRACE™ 1300 Series GC results in an extremely fast, easy to use, compact GC, delivering an incredibly high lab productivity at reduced cost of ownership.

    Product Information PDF

    Trace 1300 Series Catalogue