Thermo Scientific Dionex AutoTrace 280 Solid-Phase Extraction Instrument
The Thermo Scientific™ Dionex™ AutoTrace™ 280 Solid-Phase Extraction instrument handles large volume Solid-Phase Extraction (SPE) of liquid samples from 20 mL to 4 liters. The instrument can be configured with 1, 3, or 6 mL SPE cartridges or 47 mm solid-phase disks. The instrument automatically loads liquid samples on the SPE material then automatically extracts the analytes of interests using up to five different solvents.
Gas Regulator and Gas Gauge Range | Output: 0–30 psi (0–1.4 bars), Input: 100 psi (6.9 bars) maximum |
Net Weight | 150 lbs. (68.1 kg) |
Dimensions (h × w × d) | 23 × 25 × 27 in (57 × 63.5 × 69 cm) Sample Rack: 8 × 16.5 × 13 in) |
500 & 600 Series EPA Methods | 24 methods stored in the Dionex AutoTrace 280 SPE instrument |
Electrical | Voltage: 100, 120, 220, or 240 V ± 10% Frequency: 47–63 Hz Power: 150 Volt AC |
Liquid Management | Air Syringe: One 10 mL air syringe LH Syringe: One 10 mL liquid handling syringe 12-port Valve: Rotary, sliding Rulon® seal Valves: 3-way, Teflon® Nozzles: Stainless steel Tubing: TFE, 1/16” ID, (6 total supplied) |
Sample Pumps | Displacement: Positive Accuracy: ± 2.5% Tube Fitting: Kynar® Piston and Liner: Ceramic Non Use: Not for use with acetic acid and acetone |